National Instruments NI PXI-5422 Manual de usuario Pagina 57

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 63
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 56
© National Instruments Corporation 57 NI PXI-5422 Calibration Procedure
vi: The session handle returned from niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_OUTPUT_ENABLED
value: VI_TRUE
11. Call niFgen_Commit (niFgen Commit VI) to commit the attribute
values to the device using the following parameter:
vi: The session handle returned from
niFgen_InitExtCal
12. Wait 500 ms for the output to settle.
13. Call
niFgen_SetAttributeViInt32 to set the calibration ADC
input (niFgen property node: Calibration»Cal ADC Input) using
the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "" (empty string)
attributeID:
NIFGEN_ATTR_CAL_ADC_INPUT
value: NIFGEN_VAL_ANALOG_OUTPUT
14. Call niFgen_WriteBinary16AnalogStaticValue (niFgen Write
Binary 16 Analog Static Value VI) to set the main DAC value using
the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName:
"0"
value: 27232
15. Call niFgen_SetAttributeViBoolean to disable the analog
output (niFgen property node: Basic Operation»Output Enabled)
using the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_OUTPUT_ENABLED
value: VI_FALSE
16. Call niFgen_Commit (niFgen Commit VI) to commit the attribute
values to the device using the following parameter:
vi: The session handle returned from
niFgen_InitExtCal
17. Wait 500 ms for the output to settle.
18. Call niFgen_ReadCalADC (niFgen Read CAL ADC VI) to measure
the analog output voltage with the onboard calibration ADC using the
following parameters:
vi: The session handle returned from
niFgen_InitExtCal
numberOfReadsToAverage: 3
Vista de pagina 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 63

Comentarios a estos manuales

Sin comentarios