NI PXI-5422 Calibration Procedure 36 ni.com
3. Call niFgen_SetAttributeViInt32 to set the analog path value
(niFgen property node: Output Attributes»Analog Path) using
the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_ANALOG_PATH
value: NIFGEN_VAL_FIXED_LOW_GAIN_ANALOG_PATH
4. Call niFgen_SetAttributeViInt32 to set the gain DAC value
(niFgen property node: Calibration»Gain DAC Value) using the
following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_GAIN_DAC_VALUE
value: 2000
5. Call niFgen_SetAttributeViInt32 to set the offset DAC value
(niFgen property node: Calibration»Offset DAC Value) using the
following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_OFFSET_DAC_VALUE
value: 32767
6. Call niFgen_SetAttributeViBoolean to set the analog filter state
(niFgen property node: Output Attributes»Analog Filter Enabled)
using the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_ANALOG_FILTER_ENABLED
value: VI_FALSE
7. Call niFgen_SetAttributeViReal64 to set the pre-amplifier
attenuation (niFgen property node: Calibration»Pre-Amplifier
Attenuation) using the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_PRE_AMPLIFIER_ATTENUATION
value: 0