NI PXI-5422 Calibration Procedure 34 ni.com
Figure 4. NI PXI-5422 External Calibration Procedure
Initialize Ext Cal Session
Initialize Analog Output Calibration
Adjust Main Path Pre-Amp Offset
Adjust Main Path Pre-Amp Gain
Adjust Main Path Post-Amp
Gain and Offset
Adjust Direct Path Gain
Adjust Oscillator Frequency
Adjust Calibration ADC
Close Ext Cal Session
Initialize Calibration
ADC Calibration
Initialize Oscillator
Frequency Calibration