National-instruments Module SCXI-1503 Manual de usuario Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 79
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
Chapter 2 Connecting Signals
© National Instruments Corporation 2-5 SCXI-1503 User Manual
3-Wire Resistive Sensor Configuration
If you are using a 3-wire resistive sensor, you can reduce the error voltage
by one-half over the 2-wire measurement by connecting the device as
shown in Figure 2-3. In this configuration, very little current flows through
R
L3
and therefore R
L2
is the only lead resistance that introduces an error into
the measurement. The resulting measurement error is:
Figure 2-3. 3-Wire Resistive Sensor Configuration
VE R
L2
I
EX
=
+
SCXI-1503
Channel X
IEX+
AI+
AI–
IEX–
SCXI-1306External Sensor
ON
CH X
I = 100 µA
R
L2
R
L1
R
L3
R
T
Vista de pagina 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 78 79

Comentarios a estos manuales

Sin comentarios