National-instruments 24-Bit Half/Full-Bridge Analog Input Module NI 923 Manual de usuario Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 36
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
NI 9237 Operating Instructions and Specifications 18 ni.com
Figure 4. Typical Passband Flatness for the NI 9237
Stopband
The filter significantly attenuates all signals above the stopband
frequency. The primary goal of the filter is to prevent aliasing.
Therefore, the stopband frequency scales precisely with the data
Frequency/Data Rate
0.50.4
0.025
0.000
–0.025
–0.050
0.30.20.10
Gain (dB)