Chapter 4 Theory of Operation
© National Instruments Corporation 4-11 SCXI-1503 User Manual
Figure 4-4. Resistance-Temperature Curve for a 2,252 Ω Thermistor
The thermistor has been used primarily for high-resolution measurements
over limited temperature ranges. However, continuing improvements in
thermistor stability, accuracy, and interchangeability have prompted
increased use of thermistors in a variety of applications.
Thermistor Measurement Circuits
This section details information about thermistor measurement circuits.
The most common technique is to use a current-source, and measure the
voltage developed across the thermistor. Figure shows the measured
voltage V
0
equals R
T
× I
EX
.
Figure 4-5. Thermistor Measurement with Constant Current Excitation
,
+
V
0
V
0
= I
EX
x R
T
R
T
Thermisto
r
I
EX