SCXI-1321 Terminal Block Installation Guide 10 ni.comAssuming a strain-gauge range with a gauge factor of GF = 2 and a quarter-bridge configuration, t
© National Instruments Corporation 11 SCXI-1321 Terminal Block Installation GuideFigure 4. Half-Bridge Nulling CircuitCH+CH+CH–EX+EX–+–+–R1R24.5 kΩ4.
SCXI-1321 Terminal Block Installation Guide 12 ni.comFigure 5. Full-Bridge Nulling CircuitShunt CalibrationShunt calibration provides a method of adj
© National Instruments Corporation 13 SCXI-1321 Terminal Block Installation Guidez is the module slot of the SCXI-1121w is the channel of the module t
SCXI-1321 Terminal Block Installation Guide 14 ni.comgauge with a gauge factor of GF = 2, Vexc = 3.3 V, and R = 120 Ω produces the following:Vchange =
© National Instruments Corporation 15 SCXI-1321 Terminal Block Installation GuideYou can read a temperature at the beginning of the test and use that
321924C-01 Jan03*321924C-01*W3 Nulling circuit of Channel 2 is enabled (factory default setting)Nulling circuit of Channel 2 is disabledW4 Nulling cir
SCXI-1321 Terminal Block Installation Guide 2 ni.comitalic Italic text denotes variables, emphasis, a cross reference, or an introduction to a key con
© National Instruments Corporation 3 SCXI-1321 Terminal Block Installation GuideConnecting the SignalsNote Refer to the Read Me First: Safety and Rad
SCXI-1321 Terminal Block Installation Guide 4 ni.comFigure 1. SCXI-1321 Parts Locator Diagram1 Strain-Relief Bar2 Strain-Relief Screws3 Safety-Ground
© National Instruments Corporation 5 SCXI-1321 Terminal Block Installation GuideFigure 2. SCXI-1321 Circuit Parts Locator DiagramInstalling the Termi
SCXI-1321 Terminal Block Installation Guide 6 ni.comNote To minimize the temperature gradient inside the terminal block, move the SCXI chassis away f
© National Instruments Corporation 7 SCXI-1321 Terminal Block Installation GuideEnvironmentalOperating temperature... 0 to 50
SCXI-1321 Terminal Block Installation Guide 8 ni.comNote Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory
© National Instruments Corporation 9 SCXI-1321 Terminal Block Installation GuideThe factory default value of all the nulling resistors on the terminal